Starting pressures the model accepts
The standard model may reject some adult blood-pressure values as outside its allowed setup range.
This is an interactive explainer of cardiovascular computer-model experiments, showing how different simulated causes of high blood pressure change blood flow and pressure waves in the arteries. The controls let you explore saved results; they do not rerun the models, and the charts are not patient measurements or camera signals.
Model-derived demonstration · no participant dataThe dashed line marks the reference value of 1. A bar above it means the model produced more kidney blood flow than the reference; a bar below means less. These ratios come from model outputs, not clinical measurements, and the two route cases are not an exact pressure match.
Waveform-shape difference (normalised RMSE) measures how different two waveform shapes are: larger values mean more difference. The stiffness settings run from left to right. These are model waveforms, not measurements from a person.
A low-pass filter removes faster signal changes above its cutoff. Lower cutoffs remove more detail; higher cutoffs retain more. The line shows the screen pass rate across the cutoff settings; it does not describe transfer from artery to skin or camera.
The standard model may reject some adult blood-pressure values as outside its allowed setup range.
Changing pressure directly and changing cardiovascular properties can produce different simulated flow patterns. In separate pressure-matched model comparisons, the routes still differed in cardiac output, resistance and regional flow.
Some settings fail to settle while more extreme settings converge. Stability does not change in a simple, predictable direction.
With default settings, the model can adjust to a pressure change as if that pressure were the new normal.
Pulse’s stock aortic inflow lacks the brief backward-flow phase after ejection that appears in the distributed reference waveform.
The resistance adjustment changes kidney flow, but does not keep its share of total flow constant as pressure changes.
In the tested build, the field called “gain” changed how quickly the reflex adjusted, not its final sensitivity.
In these model runs, compliance changes altered the simulated internal-carotid shape above the chosen noise screen. This does not show that a camera can detect it.